After its introduction in 1992, the European Data Format (EDF) became the standard for EEG and PSG (Sleep) recordings. During that time, several users pointed us at its limitations also for application in other fields such as myography, evoked potentials and cardiology. A major limitation was that EDF can only handle uninterrupted recordings. So we simply skipped that limitation but kept all other specifications of EDF intact. While maintaining EDF compatibility, we also standardized most of the labels and added a possibility to save annotations and analysis results. The result is EDF+ and can save most EEG, PSG, ECG, EMG, and Evoked Potential data that can not be saved into common hospital information systems.
Using EDF+, all signals, annotations and events that are recorded in one session using one recording system can be kept safely together in one file. EDF+ can also store events and annotations only, without any signals. This flexibility allows the user to choose an optimal mix. For instance, our sleep centre stores all on-line recorded data (signals, annotations) in one file, its hypnogram and apnea detections in another EDF+ file, the same sleep scorings but made by another technician in a third file. In Neurophysiology, the on-line obtained raw EEG traces with stimulus events from an EP investigation might be stored in one file and the averaged curves with detected latencies in a second file. In Cardiology, the raw ECG with annotations about the patients exercises can be in one file, the detected QRS parameters in another file.
EDF+ allows storage of several NON-CONTIGUOUS recordings into one file. This is the only incompatibility with EDF. All other features are EDF compatible. In fact, old EDF viewers still work and display EDF+ recordings as if they were continuous. Therefore, we recommend EDF+ files of EEG or PSG studies to be continuous if there are no good reasons for the opposite.
Because EDF+ is
very close to EDF, and equally
simple, EDF+ software can relatively easily be developed based on
available
EDF
software.
Because EDF+ is
based on EDF, you should first
read the EDF specs. Section 2.1 below
describes
how EDF+ differs from EDF. Section 2.2 describes how one of the EDF+
signals
can be specially coded to store text annotations, time, events and
stimuli.
EDF+
prescribes the following filename convention. Signals, recorded using
the
same technique and constant amplifier settings, can be stored in one
file.
Different techniques, or identical techniques but with different
amplifier
settings, must be stored in separate files. All EDF+ files must have
.edf
or .EDF as filename extension. See also section 2.3.
2.1.
EDF+ compared to EDF
A standard EDF file consists of a
header
record followed by data records. The header record identifies the
patient
and specifies the technical characteristics of the recorded signals.
The
data records contain consecutive fixed-duration epochs of the recorded
signals. A standard EDF+ file also consists of a header record
followed
by data records. The structure of these records is compatible to EDF
but
contains additional specifications. Note that for your EDF+ software to
also be EDF compatible, it should support
but not rely on these
additional specifications.
2.1.1.
The
EDF+ header
The EDF+ header record identifies the patient and specifies the
technical
characteristics of the recorded signals according to the EDF
specs, except for the first 'reserved' field (44 characters)
which
must start with 'EDF+C' if the recording is uninterrupted, thus
having
contiguous data records, i.e. the starttime of each data record
coincides
with the end (starttime + duration) of the preceding one. In this case,
the file is EDF compatible and the recording ends (number x duration)
seconds
after its startdate/time. The 'reserved' field must start with
'EDF+D'
if the recording is interrupted, so not all data records are
contiguous.
In both cases, the time must be kept in each data record as specified
in
section
2.2.4.
The only incompatibility with EDF is, that signals may be recorded
discontinuously.
Therefore, we have decided that the EDF+ 'version' field must still
read
'0 ' like in EDF. In
this way, old EDF viewers will still work and display EDF+ files
(be
they continuous or discontinuous) as continuous EDF files. EDF+
software will know the difference between continuous and discontinuous
files from the mentioned 'reserved' field.
2.1.2.
The EDF+ data records
A signal
in an EDF data record is a series of 2-byte samples, the subsequent
samples
representing subsequent integer values of that signal, sampled with
equal
time intervals. We will refer to this kind of signal as an 'ordinary
signal' from now on. EDF+ data records can (and usually do) also
contain
ordinary signals. The EDF+ data records contain the ordinary signals
according
to the EDF specs (including the size limit
of
61440), except for the fact that the data records may unconditionally
be
shorter than 1s, and subsequent data records need not form a
continuous
recording. However, as in EDF, data records that follow up in time
must also follow up in the file. The samples of an ordinary signal must
have equal sample intervals inside each data record, but the interval
to
the first sample of the next data record may be different.
For instance, in a motor nerve conduction study with a number of
stimuli,
each data record would hold the ordinary signals corresponding to one
stimulus.
In this case, the duration of a data record corresponds to the "window
size" in an ENMGEP study.
Specifying a duration makes no sense if the EDF+ file does not contain any ordinary signals, for instance in a file that only contains manual sleep scores (like in both demo scoring files that come with Polyman, and the example in
3.3). This is also true
in the extreme case in which each ordinary signal only occupies one
sample
in each data record, while the file is discontinuous (EDF+D). In those two cases, specify the 'duration of a data record' to be 0.
2.1.3.
Additional specifications
in EDF+
1. In the header, use only printable US-ASCII
characters with byte values 32..126.
2. The 'startdate' and 'starttime' fields
in the header should contain only characters 0-9, and the period (.) as
a separator, for example "02.08.51". In the 'startdate', use 1985 as a
clipping date in order to avoid the Y2K problem. So, the years
1985-1999
must be represented by yy=85-99 and the years 2000-2084 by yy=00-84.
After
2084, yy must be 'yy' and only item 4 of this paragraph defines the
date.
3. The 'local patient identification' field
must start with the subfields (subfields do not contain, but are
separated
by, spaces):
- the code by which the patient is known in
the hospital administration.
- sex (English, so F or M).
- birthdate in dd-MMM-yyyy format using the
English 3-character abbreviations of the month in capitals. 02-AUG-1951
is OK, while 2-AUG-1951 is not.
- the patients name.
Any space inside the hospital code or the
name of the patient must be replaced by a different character, for
instance
an underscore. For instance, the 'local patient identification' field
could
start with: MCH-0234567 F 02-MAY-1951 Haagse_Harry. Subfields whose
contents
are unknown, not applicable or must be made anonymous are replaced by a
single character 'X'. So, if
everything
is unknown then the 'local patient identification' field would start
with: 'X X X X'. Additional subfields may follow the ones
described
here.
4. The 'local recording identification' field
must start with the subfields (subfields do not contain, but are
separated
by, spaces):
- The text 'Startdate'.
- The startdate itself in dd-MMM-yyyy format
using the English 3-character abbreviations of the month in capitals.
- The hospital administration code of the
investigation, i.e. EEG number or PSG number.
- A code specifying the responsible
investigator
or technician.
- A code specifying the used equipment.
Any space inside any of these codes must be
replaced by a different character, for instance an underscore. The
'local
recording identification' field could contain: Startdate 02-MAR-2002
PSG-1234/2002
NN Telemetry03. Subfields whose contents are unknown, not applicable or
must be made anonymous are replaced by a single character 'X'. So, if
everything
is unknown then the 'local recording identification' field would start
with: 'Startdate X X X X'. Additional subfields may follow the ones
described
here.
5. 'Digital maximum' must be larger than
'Digital
minimum'. In case of a negative amplifier gain the corresponding
'Physical
maximum' is smaller than the 'Physical minimum'. Check item 9 on how to
apply the 'negativity upward' rule in Clinical Neurophysiology to the
physical
ordinary signal. 'Physical maximum' must differ from 'Physical
minimum'.
In case of uncalibrated signals, physical dimension is left empty (that
is 8 spaces), while 'Physical maximum' and 'Physical minimum' must
still
contain different values (this is to avoid 'division by 0' errors by
some
viewers).
6. Never use any digit grouping symbol
in numbers. Never use a comma "," for a for a decimal separator. When a
decimal separator is required, use a dot (".").
7. The ordinary signal samples (2-byte two's
complement integers) must be stored in 'little-endian' format, that is
the least significant byte first. This is the default format in PC applications.
8. The 'starttime' should be local time at
the patients location when the recording was started.
9. Use the standard texts and polarity rules
at http://www.edfplus.info/specs/edftexts.html.
These standard texts may in the future be extended with further texts,
a.o. for Sleep scorings, ENG and various evoked potentials.
10. The 'number of data records' can only
be -1 during recording. As soon as the file is closed, the correct
number
is known and must be entered.
11. If filters (such as HighPass, LowPass
or Notch) were applied to the ordinary signals then, preferably
automatically,
specify them like "HP:0.1Hz LP:75Hz N:50Hz" in the "prefiltering" field
of the header. If the file contains an analysis result, the
prefiltering
field should mention the relevant analysis parameters.
12. The "transducertype" field should specify
the applied sensor, such as "AgAgCl electrode" or "thermistor".
2.2. Annotations for text, time-keeping, events and stimuli
This section describes how one of the EDF+ signals can be specially coded to store text annotations, time, events and stimuli. In this way, annotations and events are kept in the same file as the signals that they refer to. The coding is EDF compatible in the sense that old EDF software would simply treat this 'EDF Annotations' signal as if it were a (strange-looking) ordinary signal.
2.2.1. The
'EDF Annotations' signal
EDF+ data records can (and often do) contain
ordinary signals. EDF+ introduces one other kind of signal, in which
the
values are annotations that can occur at any arbitrary point of time.
This
signal is identified by giving it (in the EDF+ header) the label 'EDF
Annotations'.
As in EDF, the 'nr of samples in each data record'
field in the header
specifies how many 2-byte integers this 'EDF Annotations' signal
occupies in each datarecord.
But instead of storing 'ordinary signal' samples, those 2-byte integers
are filled with characters. The character-bytes are stored
byte-by-byte
without changing their order. For instance, the text 'abc' is
represented
by successive byte values 97, 98 and 99 in the 'EDF Annotations'
signal.
Even if no annotations are to be kept, an EDF+ file must contain at
least one 'EDF Annotations' signal in order to specify the starttime of
each datarecord (see section
2.2.4). Of course, the label 'EDF Annotations' is not allowed for
ordinary
signals.
The 'EDF
Annotations' signal only has meaningful header fields 'label' and 'nr
of
samples in each data record'. For the sake of EDF compatibility, the
fields
'digital minimum' and 'digital maximum' must be filled with -32768 and
32767, respectively. The 'Physical maximum' and 'Physical minimum'
fields
must contain values that differ from each other. The other fields of
this
signal are filled with spaces.
2.2.2. Time-stamped Annotations
Lists (TALs) in an 'EDF Annotations' signal
Text, time-keeping, events and stimuli are
coded as text annotations in this 'EDF Annotations' signal. The
annotations
are listed in Time-stamped Annotations Lists (TALs) as follows.
Each TAL starts with a time stamp OnsetDuration
in which
and
are single bytes with value 21 and 20, respectively (unprintable ASCII
characters) and Onset as well as Duration are coded using US-ASCII
characters
with byte values 43, 45, 46 and 48-57 (the '+', '-', '.' and
'0'-'9'
characters, respectively). Onset must start with a '+' or a '-'
character
and specifies the amount of seconds by which the onset of the annotated
event follows ('+') or precedes ('-') the startdate/time of the file,
that
is specified in the header. Duration must not contain any '+' or '-'
and
specifies the duration of the annotated event in seconds. If such a
specification
is not relevant, Duration can be skipped in which case its
preceding
must also be skipped. Both Onset and Duration can contain a dot ('.')
but
only if the fraction of a second is specified (up to arbitrary
accuracy).
After the time stamp, a list of annotations all sharing the same Onset
and Duration may follow. Each annotation is followed by a single
and may not contain any .
A -byte (the
unprintable
ASCII character with byte value 0) follows after the last
of this TAL. So the TAL ends with a
followed by a .
In each data record, the first TAL must start at the first byte of the
'EDF Annotations signal'. Subsequent TALs in the same data record must
follow immediately after the trailing
of the preceding TAL. A TAL, including its trailing ,
may not overflow into another data record. Each event is annotated only
once, even if its duration makes it extend into the time period of
other
data records. Unused bytes of the 'EDF Annotations' signal in the
remainder
of the data record are also filled with -bytes.
Additional 'EDF Annotations' signals may be defined according to the
same
specification.
For example, if
the technician switches off the lights and closes the door 3 minutes
after
startdate/time, this can be stored as the 28-bytes TAL '+180Lights
offClose door'
without the quotes. Alternatively, the two events can be stored as two
separate shorter TALs '+180Lights
off+180Close
door',
also without the quotes. The TAL '+1800.225.5Apnea'
codes a 25.5s apnea that begins 30 minutes and 0.2s after starttime.
2.2.3. Annotations
in a TAL
The part between
and the next is
called
one annotation. These annotations may only contain UCS characters (ISO
10646, the 'Universal Character Set', which is identical to the Unicode
version 3+ character set) encoded by UTF-8.
This encoding is supported by the major operating systems, compilers
and
applications. The first 127 UCS characters are identical to those in
US-ASCII
and are encoded in the corresponding single byte values. US-ASCII
characters
that are represented by byte values 0-31 are allowed in the annotations
only if explicitly prescribed by this EDF+ protocol. In order to enable
multi-line texts and tables, US-ASCII characters that are represented
by
byte values 9 (TAB), 10 (LF) and 13 (CR) are allowed in the
annotations.
The first 65534 characters (the Basic Multilingual Plane: BMP) of the
UCS
contain virtually all characters used in any language in the world
including
Asian languages and UTF-8 encodes these in up to three byte-values.
Remember
that this encoding applies to the 'EDF Annotations' signal only: in the
EDF+ header, only US-ASCII characters with byte values 32..126
are
allowed.
In order to support automatic averaging and superimposition, identical
events or stimuli that occur several times in one file must be coded
each
time by the same, unique annotation. Annotations (the part between
and the next )
of
different events/stimuli (or the same stimulus at a different
location)
must differ from this unique annotation.
Annotations, for instance stimuli, that are related to information in
only
one particular data record, must be in that same data record. Even
annotations
describing events preceding the start of that data record, for instance
a pre-interval stimulus must
follow the time-keeping annotation.
2.2.4. Time keeping
of data records
Because data records need not be contiguous,
the starttime of each data record must be specified in another way. So,
the first annotation of the first 'EDF Annotations'
signal
in each data record is empty, but its timestamp specifies how many
seconds
after the filestartdate/time that data record starts. So, if the first
TAL in a data record reads '+567',
then that data record starts 567s after the startdate/time of the file.
If the data records contain 'ordinary signals', then the starttime of
each
data record must be the starttime of its signals. If there are no
'ordinary
signals', then a non-empty annotation immediately following the
time-keeping
annotation (in the same TAL) must specify what event defines the
starttime
of this data record. For example, '+3456.789R-wave
indicates that this data record starts at the occurrence of an R-wave,
which is 3456.789s after file start.
The startdate/time of a file is specified in the EDF+ header fields
'startdate
of recording' and 'starttime of recording'. These fields must indicate
the absolute second in which the start of the first data record falls.
So, the first TAL in the first data record always starts with
+0.X,
indicating that the first data record starts a fraction, X, of a second
after the startdate/time that is specified in the EDF+ header. If X=0,
then the .X may be omitted.
Ideally, all
data (signals, annotations, events)
recorded in one session using one recording system are in one EDF+
file.
Data from the same patient but from other sessions or equipment will
usually
be kept in separate files. Ideally, all these files have an identical
'Patient
identification' field. In this way, accurate synchronicity between
signals
and events is kept within the files and it is exactly known to what
period
in which patients life the data apply.
In practice, this will not always be possible. However, it ķs
easy
to maintain synchronicity and patient identification between a
recording
and data that are derived from that recording. Such derived data can be
analysis results such as averages, QRS parameters, peak latencies or
sleep
stages or simply a subset of the recording. If such analysis results
are
stored in EDF+ then this must be done as follows.
If the original recording is in file R.edf (R can be any string), then
the derived-file name must be RA.edf in which A can be any string. For
instance a PSG would be recorded in file PSG0123_2002.edf and its sleep
stage analysis in PSG0123_2002_hyp.edf. Copy the patient-id line
(80
characters) from the recorded file into the analysis file.
Make sure that startdate, starttime, and number and duration of
datarecords,
are correct. So, if the analysis contains a period from 01:05:00 till
01:25:00
of a 24-hour recording that was started on August 2, 1999, 23:00:00hr,
then the analysis file should have startdate 03.08.99 and starttime
01.05.00.
In this way it is clear that both files refer to one and the same time
period in the patient's life. Some viewers (like PolyMan) are then
capable
of showing the two (or more) files time-synchronized on one screen.
Because
the analysis may reduce or increase the amount of data, the durations
of
analysis-file data records and recording-file data records may differ.
Apply suitable scaling factors in such a way that a large part of the
available
range of -32768 till 32767 for the values of the analysis results is
used.
If necessary, the scaling factor can be adapted to the dynamic range of
the analysis result even after the analysis was done. Put these scaling
factors in the header (digital and physical minimum and maximum) of the
analysis file. If such scaling is really impossible because the useful
dynamic range of the analysis result is too large, but only then, apply
the standardized logarithmic transformation
to store floating point values. Be aware that old EDF software is not
aware
of this transformation, and will show the analysis results on a
logarithmic
scale. So really try scaling first!
If a hypnogram is stored as an ordinary signal, sleep stages
W,1,2,3,4,R,M
should be coded in the data records as the integer numbers
0,1,2,3,4,5,6
respectively. Unscored epochs should be coded as the integer number 9.
If a hypnogram is stored as annotations, use the standard
texts.
Automatically document the analysis principle and parameters in the
Recording-id
and, in case of ordinary signals, also in Label, Transducer type,
Physical
dimension and Prefiltering fields in the header of the analysis file.
+0Stimulus
click 35dB both earsFree
text -0.065Pre-stimulus beep 1000Hz |
+0.3Stimulus
click 35dB both ears +0.235Pre-stimulus beep 1000Hz |
3.2. Sleep recording (PSG) with MSLT
A PSG that is followed by an MSLT can be stored in separate files. The
PSG file, including lights-off and
final
wake-up annotations, is a continuous EDF+ file. The MSLT is a
discontinuous EDF+ file which contains only the
20-minute
periods in bed. Alternatively, the PSG and the MSLT can also be stored
together into one single (discontinuous) file.
+0Recording
starts +0660Sleep stage W +120Lights off +660300Sleep stage N1 +742Turning from right side on back +960180Sleep stage N2 +993.21.2Limb movementR+L leg +1019.40.8Limb movementR leg +1140300Sleep stage N3 +1526.830.0Obstructive apnea +1603.224.1Obstructive apnea +1440210Sleep stage N2 +1650270Sleep stage N3 +1634Turning from back on left side +192030Sleep stage N2 . . . . . . . . . . . . . . . . . . +30100Lights on +30210Recording ends |
3.4. A large neurophysiological session
The
example session includes the following investigations that are stored
in separate
EDF+
files. Note that the EDF+ processing software must implement any
'negativity
upward' rule after reading the signal from the file. See further label
and polarity rules at http://www.edfplus.info/specs/edftexts.html.
A
continuous EMG
is stored in a file with two
signals:
the raw EMG and the obligatory 'EDF Annotations'
signal.
This is a 'continuous' EDF+ file, so the recording might alternatively
be stored as an EDF file. In case of a concentric needle electrode
recording,
a positivity at the centrally insulated wire relative to the cannula of
the needle is stored as a positive value in the file.
An
F response
is also stored in a file with the raw EMG and the annotations. The
duration of a data
record equals the duration of the investigator's screen, i.e. of the
'window'
(e.g. 50 ms). Each data
record contains one single response. The annotations describe timing
and any further characteristics of the
stimulus.
They can also describe measured distances and latencies.
A
Motor
Nerve Conduction Velocity with one EMG channel is also stored in a file with the raw EMG and the
annotations.
The duration of a data record equals the duration of the 'window'. The
curves from wrist and elbow stimulation are stored in the first and the
second data record, respectively. The annotations describe
timing and further characteristics of the stimulus. They can also
describe measured
distances
and latencies. This file is described in full detail below.
A
Somatosensory Evoked
Potential with four recorded signals is stored in a file with five
signals: the four raw SSEP signals and one 'EDF
Annotations'
signal. The duration of a data record equals the duration of the window
(e.g. 100 ms). The annotations describe timing and characteristics of
the
stimulus. Another EDF+ file contains
the 4-channel averaged responses (averages of odd and even sweeps are
kept
in different data records) and the 'EDF Annotations' signal which
stores
stimulus characteristics and measured latencies.
A
Visual Evoked Potential
is investigated by recording two sagittal EEG signals during
checkerboard
stimulation of the left and right field, respectively.
The left- and right-stimulated averages are
stored in two separate files.
Left and right investigations are repeated once for checking
reproducibility,
so both files contain two data records. Each data record lasts 300ms
and
contains three signals: two EEG averages and the EDF Annotations. The
annotations
describe stimulus characteristics and measured latencies. The signal
sampling
starts 10ms before each stimulus, so the first two TALs in the 'left'
file
is 0.000
and 0.010Stimulus
checkerboard left.
3.5. Intra-operative monitoring
Four
(left and right) signals with alternating right and left stimulus are
monitored. In
this
case, left and right stimulation affect correspondingly lateralized
signals.
So, the recording can be stored in two ('left' and 'right') EDF+ files.
Each file contains 4 electrophysiological signals and 1 'EDF
Annotations'
signal. Alternatively, if sufficient amplifiers are available, the
recording
can be stored in one file containing 9
signals
(4 'left' and 4 'right' electrophysiological signals and 1 'EDF
Annotations'
signal). In both cases, the response to each stimulus is stored in a
separate
data record and the 'EDF Annotations' signal specifies timing and
characteristics
of the stimulus (a.o. whether it is left or right).
3.6. Routine EEG
The 10/20 system electrodes (for instance F3, C3, T3, Cz
and O1 and so on) are recorded against
a common
reference and saved as such in the EDF+ file. Therefore the montages can be made during review, such
as F3-C3, T3-C3, C3-Cz and C3-O1. Because
electrode
locations are specified using standard
texts, re-montaging (i.e. re-referencing) EEG derivations can be
done
automatically. The Annotations signal contains events such as 'Eyes Closed' or 'Hyperventilation'.
3.7. The Motor Nerve Conduction file
A right Median Nerve conduction velocity is investigated by recording the right Abductor Pollicis Brevis while stimuli are given at wrist and elbow. The averaged signal and the corresponding annotations are stored in two data records: one for wrist and the other for elbow stimulation.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1st signal |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Each data record
contains:
1000 * 2-byte integer : R APB samples
60 * 2-byte integer : EDF Annotations
+0Stimulus right wrist 0.2ms x 8.2mA at 6.5cm from recording siteResponse 7.2mV at 3.8ms |
+10Stimulus right elbow 0.2ms x 15.3mA at 28.5cm from recording siteResponse 7.2mV at 7.8ms (55.0m/s) |
In this example, the TALs take less than 100 characters per data record, so the header reserves 120 characters (60 'samples') for the EDF Annotations signal.
+10 |
+10Stimulus_elbow |
+10<EDF_XMLnote> <Stimulus_elbow><duration unit="ms" >0.2</duration> <intensity mode="current" unit="mA">15.3</intensity> <position>right elbow</position> <distance mode="stimulus to recording" unit="cm">28.5</distance> </Stimulus_elbow> </EDF_XMLnote> |
+10<EDF_XMLnote> <measurements> <latency unit="ms">7.8</latency> <amplitude mode=" baseline to peak" unit="mV">7.2</amplitude> <velocity mode = "segmental" unit = "m/s">55.0</velocity> </measurements> </EDF_XMLnote> |